Why it shows“Not all channels support on-demand operations. Only clocked operations are allowed”when I use NI-DAQ to acquire data?
34 visualizaciones (últimos 30 días)
Mostrar comentarios más antiguos
Hi, all,
When I want to use MATLAB (R2022b-academic use) to control NI-USB 4431 to acquire data, the command window shows "Not all channels support on-demand operations. Only clocked operations are allowed." What does this mean? I can't figure out how to solve it (I have installed NI-DAQmx toolbox, version 22.2.0). Here is the code:
dq = daq("ni");
dq.Rate = 8000;
ch1 = addinput(dq, d.DeviceID, "ai1", "Voltage");
tabledata = read(dq)
In addition, the NI-4431 device information is as follows.
ni: National Instruments(TM) USB-4431 (Device ID: 'Dev1')
Analog input supports:
-10 to +10 Volts range
Rates from 1000.0 to 102400.0 scansc
4 channels ('ai0','ai1','ai2','ai3')
'Voltage','Accelerometer','Microphone','IEPE' measurement types
Analog output supports:
-3.5 to +3.5 Volts range
Rates from 800.0 to 96000.0 scansc
1 channel ('ao0')
'Voltage' measurement type
Thanks!
0 comentarios
Respuestas (1)
Manikanta Aditya
el 20 de Mzo. de 2023
Hi Hunag,
As per my understanding, you are getting error saying ‘Not all channels support on-demand operations. Only clocked operations are allowed.’ When trying to use NI-DAQ to acquire data.
The specific input channel you are trying to use does not support on-demand operations, which is when data is acquired on demand as opposed to at a regular interval. Instead, it only supports clocked operations, which require you to specify a regular clock signal to trigger data acquisition.
As a workaround, you can add a clock signal to your data acquisition session by using the "addclock" function.
dq = daq("ni");
dq.Rate = 8000;
ch1 = addinput(dq,"Dev1","ai1","Voltage");
clock = addclock(dq,"ScanClock","ai1",dq.Rate);
start(dq,"Duration",1);
tabledata = read(dq)
For further reference, please check out this link to know more about ‘addclock’ function:
I hope this resolves the issue you are facing.
1 comentario
KOLBY
el 2 de Jul. de 2024
Hello,
I am facing a similar issue when using a MCC USB 205 DAQ board. I receive "Added channel does not support clocked sampling: clocked operations are disabled. Only on-demand operations are allowed. " after adding an output channel. Additionally, this sets the Rate of the board to 0, preventing me from implementing a clock.
While this is the opposite warning, do you know how to use simultaneous analog input and output given the error received?
Is this a limitation of the board, or is there a workaround in MATLAB?
Thank you,
KS
Code used to troubleshoot, error follows that readwrite is limited to clocked operations.
clear; clc; close all;
d = daq("mcc")
addoutput(d, "Board1","ao1","Voltage");
d
addinput(d, "Board1","ai0","Voltage");
d
% clock = addclock(d,"ScanClock","Board1/ai0",d.Rate);
outScanData = linspace(0,1,d.Rate)'; % Increase output voltage with each scan.
inScanData = readwrite(d,linspace(0,1,d.Rate)');
Ver también
Categorías
Más información sobre Analog Data Acquisition en Help Center y File Exchange.
Community Treasure Hunt
Find the treasures in MATLAB Central and discover how the community can help you!
Start Hunting!