visadev - how to know bytes available

20 visualizaciones (últimos 30 días)
covfefe
covfefe el 29 de Ag. de 2025 a las 20:54
Comentada: covfefe el 30 de Ag. de 2025 a las 5:32
Hello,
I have existing code that has been working great on 2016-2023 Matlabs:
Instrument = visa('ni', 'GPIB0::18::INSTR');
fprintf(Instrument, ['MMEMory:STORe:SCReen ', DeviceFile]);
fprintf(Instrument, ['MMEMory:DATA? ', DeviceFile]);
[DataRead, count] = fread(Instrument.HANDLE);
When I updated to the latest Matlab, the fread now takes extremely long (times out), giving the following warning:
transportlib:client:ReadWarning
Warning: The specified amount of data was not returned within the Timeout period for 'read'.
'visadev' unable to read any data. For more information on possible reasons, see visadev Read Warnings.
It seems fread now wants a 2nd argument, specifying the number of Bytes to read. But this does not seem posible to get:
  • The instrument (Agilent E4443A) does not have a command to provide the size of the file
  • Instrument.NumBytesAvailable has been depreacted
What is the best way to read an unknown Byte amount with visadev (similar to previous gpib/visa implementations):
  • Can I read until a terminator?
  • Can I check the number of Bytes available on the buffer?
thank you

Respuestas (1)

Walter Roberson
Walter Roberson el 30 de Ag. de 2025 a las 4:17
You can configureTerminator followed by readline
If your data is binary then you could uint8(char()) the readline() string result and then typecast to the actual datatype.
If your data is binary then you could potentially write it in "binblock" format and use readbinblock
There does not appear to be any supported way to determine the size of the input queue. (I do not have any hardware that supports VISA so I cannot test for unsupported ways.)
  1 comentario
covfefe
covfefe el 30 de Ag. de 2025 a las 5:32
I just tried all 3 possible values for the Terminator, without luck:
K>> configureTerminator(Instrument.HANDLE,"CR")
K>> GPIBwrite(Instrument, ['MMEMory:DATA? ', DeviceFile]); [DataRead, count] = fread(Instrument.HANDLE);
Warning: The specified amount of data was not returned within the Timeout period for 'read'.
'visadev' unable to read all requested data. For more information on possible reasons, see visadev Read Warnings.
K>> configureTerminator(Instrument.HANDLE,"LF")
K>> GPIBwrite(Instrument, ['MMEMory:DATA? ', DeviceFile]); [DataRead, count] = fread(Instrument.HANDLE);
Warning: The specified amount of data was not returned within the Timeout period for 'read'.
'visadev' unable to read all requested data. For more information on possible reasons, see visadev Read Warnings.
K>> configureTerminator(Instrument.HANDLE,"CR/LF")
K>> GPIBwrite(Instrument, ['MMEMory:DATA? ', DeviceFile]); [DataRead, count] = fread(Instrument.HANDLE);
Warning: The specified amount of data was not returned within the Timeout period for 'read'.
'visadev' unable to read all requested data. For more information on possible reasons, see visadev Read Warnings.

Iniciar sesión para comentar.